Hospital Defibrillator with Patient Monitor Comen S5
S5 biphasic defibrillator/monitor +NIBP/SPO2/PR/EtCO2/PACER with 5000mAh battery (standard edition)
S5 is a biphasic defibrillator/monitor, and is applicable to manual defibrillation, AED defibrillation, pace-making and vital signs monitoring on patients.
The key features including, 7 inch LCD color display with LED backlight
Can be operated through buttons, keys and rotary knob
Up to 360J energy selection
Integrated vital signs monitoring: ECG, NIBP, SpO2, EtCO2, PR.
Rechargeable battery supporting long working time
Powerful data storage
Built-in printer
Specifications:
Size: 295mm?252mm?316mm
Weight 5.6 Kg (Including 1 battery), 5.384 Kg (Main unit)
Screen Size: 7” TFT screen
Resolution 800 x 480
Waveforms: Max 4 waveforms
Operation Environment
Temperature: 0~45 oC
Humidity:
Atmosphere Pressure:
10%~95%, non-condensation
700hPa~1060hPa
Ingress Protection: IP44
Power requirement: 100-240V~, 50/60Hz±3Hz
Battery type: Rechargeable Lithium-ion battery
Battery capacity:
7500mAh, d.c.14.8V
5000mAh, d.c.14.8V
Battery number: 1
Battery recharging
Time:
7500mAh Battery: Less than 2
hours to 80% and less than 3 hours to 100% with equipment power off
5000mAh Battery: Less than 1.5 hours to 80% and less than 2.5hours to 100% with equipment power off
Battery backup:
7500mAh Battery:
Monitoring Mode: no less than 6 hours
Defib Mode: 210 times (360J charge at intervals of 1minute without recording),
Pacing Mode: 4.5 hours (Load:50 Ω, frequency: 80bpm, current: 60mA, without recording)
5000mAh Battery:
Monitoring Mode: no less than 4 hours
Defib Mode: 120times (360J charge at intervals of 1minute without recording),
Pacing Mode: 3hours (Load:50Ω, frequency: 80bpm, current: 60mA, without recording)
Brightness: Manual from X to 100, X refers to the darkest brightness (X is 10 by default)
Indicator
Two alarm indicators
Power indicator
Battery indicator
Maintain indicator
QRS beep and alarm sound
Operating key sound
Interfacing
USB interface
RJ45 interface
AC power input
Multi-functional connector
Date storage
Alarm Event: 200 groups
Patient profiles: 100 groups
Patient Events: 1000 groups
Wave Review: 10min
NIBP Review: 2000 groups
Trend Graph: 160 hours
Trend Table: 160 hours
Voice recording: Max 240 min in total (Up to 60 min for each patient)
Marked events Available
Power-off storage: Yes
Alarm: User-adjustable High and Low 3-level Limits, Prioritized audible and visual alarm
Network: Connected to Central Monitoring System by hardwire/wireless
Recorder
Type: Built-in, Thermal array
Channel: Max 3 channel waveforms
Real-time recording: 3s, 5s, 8s, 16s, 32s, Continual
Speed: 6.25mm/s, 12.5mm/s, 25mm/s, 50mm/s
Record width: 50mm
Resolution: 8dot/mm (Horizontal and vertical)
Background grid: Configurable
External printer: Yes
Defibrillation
Operating mode: Manual Mode, AED Mode, Synchronous Defibrillation
Waveform: Biphasic truncated exponential waveform, with impedance compensation
Defibrillation pathway: External defibrillation
Electrode type: External defibrillation paddles, multifunctional electrode External defibrillation electrode paddles:
Supports charging, discharging and energy selection, Charging completion indicator
Charge Time (Battery power):
Less than 3 seconds to 200 Joules with a new, fully charged battery
Less than7 seconds to 360 Joules with a new, fully charged battery
Charge Time (AC power):
Less than4 seconds to 200 Joules
Less than 8 seconds to 360 Joules
Energy accuracy: ±1.5J or ±10% of setting, whichever is greater, while 50 Ω impedance ±2J or 15% of setting, whichever is greater, while 25Ω, 75Ω, 100Ω, 125Ω, 150Ω, 175Ω impedance
Patient Impedance
Range:
20~300Ω(External defibrillation),
Defibrillation proof: Type CF: ECG, RESP, SpO2, NIBP, PR
Type BF: EtCO2
Manual Mode
External defibrillators: 1J~360J, 25 types (1/2/3/4/5/6/7/8/9/10/15/20/30/50/70/100/120/150/170/200/220/250/270/300/360J)
Synchronous
Cardioversion:
Energy transfer begins within 60ms of the R wave from internal Sync signal
Energy transfer begins within 25ms of the External Sync signal
AED
Output Energy: Adujustable:100-360J
Number of electric shocks
Adjustable: once, twice, 3 times
Types can be AED: VF & VT
AED maximum time required for cardiac rhythm analysis to be ready for discharge:
Battery power supply: 18s
AC power supply: 21s
Noninvasive Pacing
Waveform: Monophasic square wave pulse
Pulse Width: 20ms or 40ms
Accuracy: ±5%
Pacing Mode: On-demand or fixed
Pacing frequency: 30 ppm to 210 ppm
Accuracy: ±1ppm or ±1.5% (whichever is greater)
Pacing output: 0 mA to 200 mA
Accuracy: ±5% or ±5mA, whichever is greater
Speed-down pacing: pacing pulse frequency reduced to 25% of original value
Monitoring
ECG (leads)
Lead Type: 3 leads ECG, 5 leads ECG, AUTO
Lead selection: 5-lead: I; II; III; aVR; aVL; aVF; V
3-lead: I; II; III
Multi-lead
synchronization analysis: Available
ECG sensitivity: Auto, 1.25 mm/mV (?0.125),
2.5 mm/mV (x0.25), 5 mm/mV (x0.5), 10 mm/mV (x1), 20 mm/mV (x2), 40 mm/mV (x4)
Accuracy: Less than ±5%
Sweep speed: 6.25 mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s
Accuracy: Less than ±10%
Heart Rate: Adult: 15~300 bpm
Pediatric: 15~350 bpm
Accuracy: ±1bpm or ±1% (whichever is greater)
Alarm limit range Adult:
High limit: (low limit+2bpm) ~ 300bpm
Low limit: 15bpm~ (highlimit-2bpm)
Pediatric:
High limit:(low limit+2bpm) ~ 350bpm
Low limit: 15bpm ~ (highlimit-2bpm)
Resolution: 1 bpm
Accuracy: ±1bpm
Bandwidth: Monitoring: 0.5~40Hz (-3.0dB~+0.4dB)
Diagnosis: 0.05~150Hz (-3.0dB~+0.4dB)
Surgery: 1~20Hz (-3.0dB~+0.4dB)
ST: 0.05~40Hz(-3.0dB~+0.4dB)
CMRR: Monitoring: >105dB
Diagnosis: >90dB
Surgery: >105dB
ST: >105dB
Input Impedance: ?5MΩ
Input signal range: ±8mV
HR trigger threshold 200μV
Lead off detection current:
Measuring electrode: <0.1μV
Driving electrode: <1μV
Pacemaker pulse suppression switch:
Manual selection when the pacemaker is turned on
Analog output: Magnification: 1:1000
Accuracy: ±5%
Bandwidth: 0.5Hz ~ 40Hz
Delay: ? 35ms
ST Detection: -2.0mV ~ +2.0mV (-20.0mm~+20.0mm)
Resolution: 0.01mV
Accuracy: -0.8mV ~ +0.8mV: ±0.02mV or ±10%
Others: Unspecified
ST analysis review 20 groups
System noise: Less than 25μV
Calibration voltage 1 mV; Accuracy: ±5%
Arrhythmia Analysis: 26 Types
Pacemaker detection: Detectable ECG (paddle)
Lead Type: Single lead ECG
Heart Rate measurement & alarm range:
Adult: 15~300bpm
Pediatric:15~350bpm
Resolution: 1 bpm
Accuracy: ±1% or ±1bpm (whichever is greater)
Bandwidth: Defib: 1~20Hz (-3dB~+0.4dB)
CMRR: Defib: >105dB
Input Impedance: ?5MΩ
Input signal range: ±8mV
HR trigger value 200μV
Arrhythmia Analysis: 5 Types, ASY, VF, VT, PNC, and PNP Respiration
Method: Thoracic Impedance Method RR measurement range:
Adult: 0~120bpm
Pediatric: 0 ~150bpm
Accuracy: 7~150bpm: ±2bpm or ±2% (whichever is greater)
0~6bpm: unspecified
Apnea Alarm: Adult: 10s~60s Ped: 10s~40s
Accuracy: ±5s
Alarm: Audible and visual alarm; alarm events reviewable COMEN NIBP
Method Automatic oscillometric
Work mode: Manual / Automatic/Continuous
Interval Time: Adjustable 1/2/2.5/3/4/5/10/15/30/60/90/120/180/240/480/720 min Continuous: 5min
Maximum measurement cycle Adu/Ped: 120s
Measurement Unit: mmHg / kPa selectable
Pressure types: Systolic, Diastolic, Mean