The availability at retail stores may vary. For better service, order online or contact the store
Click on an image to open Gallery
The availability at retail stores may vary. For better service, order online or contact the store
HANNA® offers you a choice of 2 meters to meet your exacting requirements. HI 991300 and HI 991301 offer you the combination of pH, electro-conductivity, total dissolved solids and temperature. To increase precision, you can select the meter which will work best with your range of conductivity, from purified to brackish waters.
There are only 2 buttons, yet you can select from a range of calibration buffers and even the temperature scale (°C or °F) most familiar to you. The housing is waterproof and rated for IP 67 conditions.
The multi-parameter probe, HI 1288, includes pH, EC/TDS and temperature in one convenient, rugged handle.
Other user selectable features include different TDS factors from 0.45 to 1.00, and a range of temperature coefficients (ß) from 0.0 to 2.4% for greater consistency and reproducibility. Also selectable are standardized buffer recognition values.
To ensure against interference from transient electrical noise, a solid-state amplifier is integrated into the probe.
Range | pH | 0.00 to 14.00 pH |
Range | EC | 0.00 to 20.00 mS/cm |
Range | TDS | 0.00 to 10.00 ppt (g/L) |
Range | Temperature | 0.0 to 60.0°C (32.0 to 140.0°F) |
Resolution | pH | 0.01 pH |
Resolution | EC | 0.01 mS/cm |
Resolution | TDS | 0.01 ppt |
Resolution | Temperature | 0.1°C (0.1°F) |
Accuracy | pH | ±0.01 pH |
Accuracy | EC | ±2% F.S. |
Accuracy | TDS | ±2% F.S. |
Accuracy | Temperature | ±0.5°C (±1.0°F) |
TDS Conversion Factor | selectable 0.45 to 1.00 with 0.01 increments (default 0.50) | |
pH Calibration |
automatic at 1 or 2 points with 2 sets of memorized standard (pH 4.01/7.01/10.01 or 4.01/6.86/9.18) |
|
EC/TDS Calibration |
automatic, 1 point at 6.44 ppt (CONV 0.5) at 9.02 ppt (CONV 0.7) at 12880 µS/cm (others) |
|
Temperature Compensation | pH | Automatic |
Temperature Compensation | EC/TDS | Automatic with selectable ß from 0.0 to 2.4%/°C with 0.1 increments |
Probe | HI 1288, pH/EC/TDS/T, DIN connector, 1 m (3.3') cable (included) | |
Battery Type / Life | 4 x 1.5V AAA / approx. 500 hours of continuous use | |
Environment | 0 to 50°C (32 to 122°F); RH max 100% | |
Dimensions | 150 x 80 x 36 mm (5.9 x 3.1 x 1.4") | |
Weight | 210 g (7.4 oz.) |
Having trouble with any of our products or just want to return? You can return it within 14 days of receipt or ask for replacemend or refund.
Request an official quote for the products you have collected in the cart with one click!
Over 70 years now we are active in the field of health, with high quality products and services, from the largest domestic and international health product houses!